|
1
|
- The spectrophotometer is an analytical instrument that makes possible a
quantitative measurement of the light passing though a solution.
- The spectrophotometer is also known as a colorimeter or a spec-20.
- This instrument has a diffraction grating monochromator, an electronic
detection, amplification and measuring system.
- It operates in a wavelength range
of 390 – 800 mu.
|
|
2
|
- White light emanating from the tungsten lamp of the Spec-20 passes
through an entrance slit and is dispersed by a diffraction grating.
- From the dispersed beam, a narrow band of a selected wavelength is
passed through a second slit and into the sample solution being
measured.
|
|
3
|
- Any of this light which is not absorbed by the solution, but which
passes through it, falls upon the phototube of the instrument where the
intensity of the transmitted light is measured electronically.
|
|
4
|
- The diffraction grating is a precision
grating having 600 grooves to the millimeter and they are
accurately spaced.
- The white light falling upon the grating is dispersed into a horizontal
fan of beams (violet and ultra-violet) at one end and the long
wavelengths (red and infrared) at the other.
|
|
5
|
- The whole spectrum of light falls
on a dark screen with a slit cut in it.
- Only that portion of the spectrum (wavelength) which happens to fall on
the slit goes through into the sample, and we can project any part of
the spectrum onto the slit that we wish simply by turning the grating.
|
|
6
|
- The grating is turned by the knob on the top of the instrument (the
wave-length control knob).
- Attached to this knob is a dial
calibrated in wavelength.
- This dial may be set to the wavelength wanted.
|
|
7
|
- The wavelengths are given in
millimicrons [1mu = 10A = 10-7 cm].
- The slit of the instrument passes
a band of wavelength 20 millimicrons wide.
|
|
8
|
- To turn on the Spec-20 rotate the
amplifier control (left-hand knob) clockwise.
- This should be done at least 20 minutes before measurements are made.
- After the instrument has warmed
up, the amplifier control knob may be adjusted so the meter needle will
read “0” on the percentage transmission scale when no light is striking
the phototube.
|
|
9
|
- The right-hand knob regulates the
amount of light passing through the second slit of the phototube.
- After the colorimeter has been
“zeroed” (by means of the amplifier control knob), a “blank” solution is placed in the
light path and the light control knob is rotated until the dial reads
100 %T.
|
|
10
|
- The “blank” solution (the medium in which the substance being measured
is located) may absorb light of certain wavelengths and interfere with
the light being absorbed.
- We will use distill water as the blank because all of our solutions have
water as the solvent.
|
|
11
|
- If a sample solution is now placed in the light path, any change in the
%T reading is due to the particular light-absorbing species in the
sample and the %T reading is a measure of the quantity of that species
present.
|
|
12
|
- Of course, whenever a change in wavelength is made, however slight, the
0 %T and 100 %T must then be reset, since the amount of “compensation”
needed varies with wavelength.
|
|
13
|
- The special tubes the “blank” and sample are placed in to be read by the
spec-20 are called cuvettes. The
handling of the cuvettes is extremely important.
- 1. Do not handle the lower
portion of a cuvette (through which the light beam will pass).
|
|
14
|
- 2. Wipe off any liquid drops, or
smudges on the lower half of the cuvette, with a Kimwipe before placing
the cuvette in the instrument.
NEVER wipe the cuvettes with paper towel or handkerchiefs.
- 3. Many precision instruments come with matched sets of cuvettes
that have index lines on them that must be matched up when placing one
in the sample holder.
|